Trouble shooting and solution of salt spray test chamber (2)
The previous section introduced three faults of the salt spray test chamber and their corresponding solutions, and this section introduces other parts. 4. When the low water level alarm light is on Cause: The water level is too low. Treatment method: check whether the water level in the test room or saturation tank is sufficient. 5. There is a normal spray and the air compressor is not running Cause of failure: The air compressor itself has a self-protection function. Treatment method: Use as usual, without affecting the test process. Six, unable to spray cause of issue: 1. The air compressor is not running; 2. The main switch of the air compressor outlet is not turned on; 3. Solenoid valve failure; 4. The pressure gauge is faulty or the pressure is too low; 5. Electromagnetic contactor failure; 6. The nozzle is blocked. Approach: 1. Turn on the air compressor button; 2. Turn on the air compressor main switch; 3.4.5. Notify the equipment manufacturer to come home for repair. 6. Remove the nozzle for cleaning (please disassemble carefully). 7. Cannot run after turning on the power Cause of failure: When the water level in the heating tank is too low, the power supply for operation will be cut off. Treatment method: increase the water level of the heating tank to normal conditions. Role Playing Game Dice Sets,Game Dice Sets,Dice Set,Games Party Dice NINGBO BESCON GAMES AND APPLIANCES CO., LTD , https://www.bescondice.com